Fluke 1586A/2HC 120 Super-DAQ and High-Capacity Module
Features
- Measure thermocouples, PRTs, thermistors, DC voltage, DC current, and resistance
- Best-in-class temperature measurement accuracy
- PRTs ±0.005 C (using external DAQ-STAQ Multiplexer)
- Thermocouples: ±0.5 C (using High-Capacity Module and internal CJC)
- Thermistors: ±0.002 C
- Input channels: Up to 40 isolated universal inputs
- Selectable scan speed: Up to 10 channels per second
- Four modes of operation: Scan, Monitor, Measure, Digital Multimeter (DMM)
- Real-time color trending: Chart up to four channels simultaneously
- Current reversal minimizes thermoelectric effects, improving PRT/RTD measurment accuracy
- Control Fluke Calibration temperature sources such as dry-wells or micro-baths for automated calibration routines
- Records up to 20 MB of data and setup files to internal non-volatile memory or to an external USB drive.
- Transfer data to a PC using the USB drive or LAN connection and view data in Microsoft Excel
- Data security: Administrator and user profiles for protecting settings and ensuring test traceability
- Mx + B scaling and channel offset zero function
- Two independent, user-defined alarms for each channel indicate when an upper or lower range has been exceeded
- 6 1/2 digit display resolution for DC voltage, DC current, and resistance
1586A Super-DAQ Precision Temperature Scanner
The 1586A Super-DAQ is the most accurate and flexible temperature data acquisition system on the market. It scans and records temperature, DC voltage, DC current, and resistance for up to 40 input channels and scan speeds as fast as 10 channels per second. The Super-DAQ can be configured for use as a multi-channel data logger in the factory or as a precision reference thermometer for benchtop sensor calibration in the lab.
When configured with the DAQ-STAQ Multiplexer (sold separately), the Super-DAQ has the accuracy of the best benchtop reference thermometer readouts for calibration of PRTs, RTDs, thermistors, or thermocouples. Lab efficiency can be increased when the Super-DAQ is connected to a Fluke Calibration drywell or bath and Automated Sensor Test routines are run. Furthermore, paring the 1586A Super-DAQ with a high-capacity module adds the ability to measure temperature, resistance, dc voltage, and dc current for data acquisition applications.
The 1586 is ideal for a number of applications such as thermal mapping, temperature validation, process sensor calibration, and more. These applications are found in a various industries including pharmaceutical, biotechnology, food processing, aerospace, and automotive.
General Specs
- Maximum input: 50 V
- Offset voltage: < 2 µV
- 3-wire internal resistance mismatch: < 50 megohm
- Basic CJC accuracy: 0.25 Celsus
High-Capacity Module
Universal input channels designed to measure temperature, resistance, dc voltage, and dc current for data acquisition applications on the factory floor where channel count and scan speed are important.
General Specs
- Maximum input 50 V
- Offset voltage < 2
V - 3-Wire internal resistance mismatch < 50 megohm
- Basic CJC accuracy 0.6 Celsius
For factory applications, the Super-DAQ is configured with the internal High-Capacity Module. Connecting thermocouples or RTDs to input terminals can be time consuming - especially if you're using many sensors of the same type for one job, and then switching to a different sensor type for another job. The internal High-Capacity Module allows you to pre-configure multiple input modules and simply exchange one module for another, depending on your test requirements. Recall a stored test setup to make the changeover even faster. And if you prefer, you can always measure a variety of different input types at the same time in a single High-Capacity Module, including thermocouples, RTDs, voltage, resistance, or current.
What's in the Box
- 1586A/2HC 120/C Super-DAQ
- (2) High-Capacity Module